EasyManua.ls Logo

Mindray BeneView T5 - Page 204

Mindray BeneView T5
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A-12
A.10 Electrical Safety Inspection after Repair
The following table specifies test items to be performed after the equipment is repaired.
Repair with main unit not disassembled
Test items: 1, 2, 3
When neither power supply
PCBA nor patient
electrically-connected PCBA is
repaired or replaced
Test items: 1, 2, 3, 4
When power supply PCBA is
repaired or replaced
Test items: 1, 2, 3, 4, 5
When patient
electrically-connected PCBA is
repaired or replaced
Test items: 1, 2, 3, 4, 6, 7, 8
Repair with
main unit
disassembled
When both power supply PCBA
and patient
electrically-connected PCBA are
repaired or replaced
Test items: 1, 2, 3, 4, 5, 6, 7, 8

Table of Contents

Other manuals for Mindray BeneView T5