2.6 BeneLink Module 
BeneLink module allows the information (patient data, alarms, etc.) from the external device 
to be displayed, saved, recorded, printed, or calculated through a BeneView patient monitor. 
If the patient monitor is connected with the CMS or gateway, information from the external 
device can also be transmitted to the CMS or gateway. BeneLink module connects with the 
external device through an ID module, which enables the information transmission between 
the BeneLink module and the external device. BeneLink module can be connected to many 
external devices such as anesthesia machine and ventilator. 
 
The following diagram shows the structure of the BeneLink module: 
AM1808 
CPU 
module
I2S
BeneLink module interface board
Isolation 
circuit
Connector 1
Connector 4
Infrared 
commun
-ication 
board
MCU
Multiplex switch
Connector
Up serial port
Serial port of external 
device
MCU serial 
port
BeneView 
patient monitor
Infrared ray
232 Serial port
Serial port ID interface board
232 Serial port
Connector
VCC
External device
ID_STATUS
ID_STATUS
ID_READ
Serial port
ID_READ
BeneLink module interface 
board for debugging
Debugging 
serial port