3-32
The test passes when the measurements are within the following ranges:
Test items Measurements
PK-PK 54V - 66V
+Width 270μs - 330μs
Checking NMT Sensor
Tools required:
None
1. Connect the patient monitor, NMT module, and NMT accessories.
2. Select [Main Menu] [Maintenance >>] [User Maintenance >>] enter the
required password [NMT Sensor Check >>] from the patient monitor.
3. Follow the on-screen instructions to check the NMT sensor in four ways.
If sensor check completes successfully, the message “Test passed. The function of NMT
sensor is OK” is presented. If any of the four steps fails, check if the sensor is placed
correctly as instructed, and perform the sensor check again. Replace the sensor if you cannot
pass the sensor check..
NOTE
z Stop NMT measurement or calibration before starting NMT sensor check.
z Take care to handle the the NMT sensor. Avoid forcefully striking the sensor.