4-12 
 
4.6.14 Device Integration Failures 
 
Symptoms   Possible Cause  Corrective Action 
The ID adapter is not 
compatible with the 
external device. 
1. Replace the ID adapter. 
2. Upgrade the ID adapter in [Factory 
Maintenance] menu to make the ID 
adapter match the corresponding external 
device. See 3.11.1 Device Connection and 
Setup for more about the setup of the ID. 
The serial port adapting 
cable is not compatible with 
the external device. 
Replace the serial port adapting cable. 
The [Devices 
Integrated] 
window displays 
nothing after 
connection. 
Wrong software version or 
wrong protocol version of 
the external device. 
Make sure the protocol version and software 
version are supported by the BeneLink 
module. 
Generate the alarm: 
[BeneLink Comm 
Stop]. 
The application of the 
BeneLink module is 
damaged. 
Update or upgrade the application of the 
BeneLink module    with the network 
upgrading tool. 
The application of the 
BeneLink module is 
damaged. 
Update or upgrade the application of the 
BeneLink module with the network 
upgrading tool. 
The patient monitor 
has no response 
when loading the 
ID adapter. 
The kernel or the document 
system of the BeneLink 
module is damaged. 
Return the BeneLink module to factory for 
repair.