Passport 12/Passport 8 Operator’s Manual A-1
A Product Specifications
A.1 Monitor Safety Specifications
A.1.1 Classifications
The monitor is classified, according to IEC60601-1:
Type of protection against electrical shock
Class I, equipment energized from an external and internal electrical
power source.
Degree of protection against electrical shock
Type BF defibrillation proof for CO
2
and AG monitoring.
Type CF defibrillation proof for ECG, RESP, TEMP, SpO
2
, NIBP, IBP and C.O..
Mode of operation
Continuous
Degree of protection against harmful ingress
of water
IPX1 (Protection against vertically falling water drops)
A.1.2 Environmental Specifications
CAUTION
The equipment may not meet the performance specifications if stored or used outside the specified
temperature and humidity ranges.
Main unit
Item Operating conditions Storage conditions
Temperature (°C) 0 to 40
Passport 8: -30 to 70
Passport 12: -20 to 60
Relative humidity (noncondensing) 15% to 95% 10% to 95%
Barometric (kPa) 57.0 to 107.4 16.0 to 107.4
Microstream CO
2
module
Item Operating conditions Storage conditions
Temperature (°C) 0 to 40 -20 to 60
Relative humidity (noncondensing) 15% to 95% 10% to 95%
Barometric (kPa) 57.3 to 105.3 57.3 to 105.3
Sidestream CO
2
module
Item Operating conditions Storage conditions
Temperature (°C) 5 to 40 -20 to 60
Relative humidity (noncondensing) 15% to 95% 10% to 95%
Barometric (kPa) 57.3 to 105.3 57.3 to 105.3