3-12 Passport 12/Passport 8 Operator’s Manual
The monitor does the following after activating Privacy mode:
The screen turns blank and the message [Under monitoring. Press any key to exit the privacy mode.] displays.
Monitoring and data storing continue but patient data is only visible at the CMS.
Alarms can still be triggered. But all audible alarms are suppressed and the alarm light is deactivated at the
monitor.
All system sounds are suppressed, including heart beat tone, pulse tone, all prompt tones, etc.
To cancel the Privacy mode, press any key.
The monitor exits the Privacy mode automatically in any of the following situations:
The monitor disconnects from the CMS.
The alarm [Battery Too Low] or [System will shut down soon. Please replace the batteries or use the
external power.] is presented.
The touchscreen is locked automatically in the Privacy mode.
WARNING
In Privacy Mode, all audible alarms are suppressed and the alarm light is deactivated at the monitor. Alarms
sound only at the CMS.
NOTE
The Privacy mode can be used only when the monitor is connected to a CMS.
3.13.4 Standby Mode
In Standby mode, you can temporarily stop patient monitoring without turning off the monitor. To enter the Standby
mode, select the [Standby] QuickKey .