Passport 12/Passport 8 Operator’s Manual C-1
C Default Configurations
This chapter lists some of the most important factory default settings for each department in configuration
management. You cannot change the factory default configuration itself. However, you can make changes to the
settings from the factory default configuration and then save the changed configuration as a user configuration.
C.1 Parameters Configuration
C.1.1 ECG
ECG Setup
Item Name
Configurable
Default
In Config Mode In Monitor Mode
Alm Source Yes Yes HR
Alarm Yes Yes On
Alm Lev Yes Yes Med
HR/PR High Yes Yes
Adu: 120
Ped: 160
Neo: 200
HR/PR Low
Yes Yes
Adu: 50
Ped: 75
Neo: 100
Gain Yes Yes X1
Sweep Yes Yes 25 mm/s
Filter Yes Yes
General: Monitor
OR: Surgery
ICU, NICU: Monitor
CCU: Diagnostic
QRS Volume Yes Yes
General, OR: 2
ICU, NICU, CCU: 1
QRS threhold No Yes 0.16 mV
Lead Set Yes Yes
Auto (if auto lead detection is available);
5-lead (if auto lead detection is not
available)
ECG Display Yes Yes Normal
Notch Filter Yes Yes Weak
Paced No Yes No
Pacer Reject No Yes Off
Smart Lead Off Yes Yes On
Pacemaker Rate (Mortara
algorithm only)
Yes Yes 60

Table of Contents

Other manuals for Mindray Passport 12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport 12 and is the answer not in the manual?

Mindray Passport 12 Specifications

General IconGeneral
BrandMindray
ModelPassport 12
CategoryMedical Equipment
LanguageEnglish

Related product manuals