C-2 Passport 12/Passport 8 Operator’s Manual
ST Analysis
Item Name
Configurable
Default
In Config Mode In Monitor Mode
ST Analysis Yes Yes General, OR, ICU, NICU: Off
CCU: On
Alarm Yes Yes Off
Alm Lev Yes Yes Med
ST-X High Yes Yes When ST Unit is mV:
When ST Unit is mm:
0.20
2.0
ST-X Low Yes Yes When ST Unit is mV:
When ST Unit is mm:
-0.20
-2.0
Auto Yes Yes Off
ISO Yes Yes -80 ms
J Yes Yes 48 ms
ST Yes Yes J + 60 ms
X represents I, II, III, aVR, aVL, and aVF.
QT/QTc Analysis
Item Name
Configurable
Default
In Config Mode In Monitor Mode
QT Analysis Yes Yes Off
QTc Formula Yes Yes Hodges
Analysis lead Yes Yes All
QTc Alarm Yes Yes Off
∆QTc Alarm Yes Yes Off
QTc Alarm Level Yes Yes Med
∆QTc Alarm Level Yes Yes Med
QTc Record Yes Yes Off
∆QTc Record Yes Yes Off
Arrh. Analysis
Arrhythmia Threshold Settings (Mindray)
Item Name
Configurable
Default
In Config Mode In Monitor Mode
PVCs High Yes Yes 10
Asys. Delay Yes Yes 5
Vtac Rate Yes Yes 130
Vtac PVC Yes Yes 6
Multif. PVC’s Window Yes Yes 15
Tachy High Yes Yes
Adu:
Ped:
120
160
Brady Low Yes Yes
Adu:
Ped:
50
75