Passport 12/Passport 8 Operator’s Manual C-3
Arrhythmia Threshold Settings (Mindray)
Item Name Configurable Default
Extreme Tachy Yes Yes
Adu:
Ped:
160
180
Extreme Brady Yes Yes
Adu:
Ped:
35
50
Vbrd Rate Yes Yes 40
Vbrd PVCs Yes Yes 5
Pause Time Yes Yes 2
Arrhythmia Threshold Settings (Mortara)
Item Name
Configurable
Default
In Config Mode In Monitor Mode
PVCs high Yes Yes 10
Asys. Delay Yes Yes 5
Vtac Rate Yes Yes 130
Vtac PVCs Yes Yes 6
Multif. PVC’s Window Yes Yes 15
Tachy High
Yes Yes Adu:
Ped:
120
160
Brady Low
Yes Yes Adu:
Ped:
50
75
Arrhythmia Alarm Settings (Mindray)
Item Name
Configurable Default
In Config Mode In Monitor Mode Alarm Switch Alarm Level
Asystole Alarm Yes Yes On High
VFib/VTac Alarm Yes Yes On High
Vtac Alarm Yes Yes On High
Vent. Brady Alarm Yes Yes On High
Extreme Tachy Alarm Yes Yes On High
Extreme Brady Alarm Yes Yes On High
PVCs/min Alarm Yes Yes
General, OR, ICU, NICU: Off
CCU: On
Med
R on T Alarm Yes Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Run PVCs Alm Lev Alarm Yes Yes Off Low
Couplet Alarm Yes Yes Off Message
Multif. PVC Alarm Yes Yes Off Med
PVC Alarm Yes Yes Off Message
Bigeminy Alarm Yes Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Trigeminy Alarm Yes Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Tachy Alarm Yes Yes Off Med