C-4 Passport 12/Passport 8 Operator’s Manual
Arrhythmia Alarm Settings (Mindray)
Item Name Configurable Default
Brady Alarm Yes Yes Off Med
Pacer Not Pacing Alarm Yes Yes Off Message
Pacer Not Capture Alarm Yes Yes Off Message
Missed Beats Alarm Yes Yes Off Message
Nonsus. Vtac Alarm Yes Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Vent. Rhythm Alarm Yes Yes Off Med
Pause Alarm Yes Yes Off Low
Irr. Rhythm Alarm Yes Yes Off Message
Afib Alarm Yes Yes Off Message
Arrhythmia Alarm Settings (Mortara)
Item Name
Configurable Default
In Config Mode In Monitor Mode Alarm Switch Alarm Level
Asystole Alarm Yes Yes On High
VFib Alarm Yes Yes On High
Vtac Alarm Yes Yes On High
PVCs/min Alarm Yes Yes
General, OR, ICU, NICU: Off
CCU: On
Med
R on T Alarm Yes Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Run PVCs Alarm Yes Yes Off Low
Couplet Alarm Yes Yes Off Message
Vent. Rhythm Alarm Yes Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Bigeminy Alarm Yes Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Trigeminy Alarm Yes Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Tachy Alarm Yes Yes Off Med
Brady Alarm Yes Yes Off Med
Pacer Not Pacing Alarm Yes Yes Off Message
Pacer Not Capture Alarm Yes Yes Off Message
Missed Beats Alarm Yes Yes Off Med
Multif. PVC Yes Yes Off Med
Irr. Rhythm Alarm Yes Yes Off Message