Passport 12/Passport 8 Operator’s Manual 7-13
7.10 Testing Alarms
When the monitor starts up, a self-test is performed. In this case the alarm lamp is lit in yellow and red respectively, and
the system gives a beep. This indicates that the visible and audible alarm indicators are functioning correctly.
For further testing of individual measurement alarms, perform the measurement on yourself (for example, SpO
2
or CO
2
)
or use a simulator. Then adjust alarm limits and check that appropriate alarm behavior is observed.
7.11 Using Care Group Alarms
7.11.1 Care Group Auto Alarms
When a Care Group is set up on your monitor, a flashing symbol will appear beside the QuickKeys area if any monitor in
your Care Group, which is not currently viewed by your monitor, is alarming. The alarm symbol is shown as below.
When a monitor in the Care Group is disconnected, the flashing symbol is shown as below.
The department and bed label of the alarming monitor appear on the symbol. You can enter the View Other Patient
window by pressing this symbol.
7.11.2 Setting Care Group Alert Tone
When a monitor in the Care Group issues an alarm, your monitor prompts you by sounding an alert tone. To set the alert
tone:
1. In the [Main Menu], select [Screen Setup >>]→[Screen Layout >>]→[Choose Screen].
2. In the [Choose Screen] window, select [View Others Screen].
3. In the view other patient window, select [Setup>>], and set [Alert Tone] to [Repeat], [Once] or [Off].
7.11.3 Resetting Care Group Alarms
You can reset the alarm of the currently viewed bed in the View Other Patient window. This function can be set in the
[Alarm Setup] menu from the [User Maintenance] menu only.
To enable this function:
1. Select [Main Menu]→[Maintenance>>]→[User Maintenance>>]→enter the required password→[Alarm
Setup>>].
2. In the [Alarm Setup] menu, set [Reset Other Bed’s Alarms] to [On].
When your monitor is viewed by other monitors and [Alarm Reset By Other Bed] is switched on, alarms on your
monitor can be reset by other monitors.

Table of Contents

Other manuals for Mindray Passport 12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport 12 and is the answer not in the manual?

Mindray Passport 12 Specifications

General IconGeneral
BrandMindray
ModelPassport 12
CategoryMedical Equipment
LanguageEnglish

Related product manuals