Passport 12/Passport 8 Operators Manual A-13
Alarm limit Range (mmHg) Step (mmHg)
Sys High
Adult: (low limit+5) to 270
Pediatric: (low limit+5) to 200
Neonate: (low limit+5) to 135
5
Sys Low 40 to (high limit-5)
Mean High
Adult: (low limit+5) to 230
Pediatric: (low limit+5) to 165
Neonate: (low limit+5) to 110
Mean Low 20 to (high limit-5)
Dia High
Adult: (low limit+5) to 210
Pediatric: (low limit+5) to 150
Neonate: (low limit+5) to 100
Dia Low 10 to (high limit-5)
**Measurement accuracy verification: In adult and pediatric modes, the blood pressure measurements measured
with this device are in compliance with the American Standard for Non-invasive sphygmomanometers (ISO
81060-2) in terms of mean error and stardard deviation by comparing with intra-arterial or auscultatory
measurements (depending on the configuration) in a typical patient population. For auscultatory reference, the
5th Korotkoff sound was used to determine the diastolic pressure.
In neonatal mode, the blood pressure measurements measured with this device are in compliance with the
American National Standard for Non-invasive sphygmomanometers (ISO 81060-2) in terms of mean error and
stardard deviation by comparing with intra-arterial measurements (depending on the configuration) in a typical
patient population.
A.7.6 Temp
Standards Meet standard of EN12470-4
Technique Thermal resistance
Measurement range 0 to 50 °C (32 to 122 °F)
Resolution 0.1 °C
Accuracy ±0.1 °C without probe)
Refreshing rate 1 s
Minimum time for accurate
measurement
Body surface: <100 s
Body cavity: <80 s
Alarm limit Range Step
T1/T2 High (low limit +1) to 50°C
0.1 °C
T1/T2 Low 0.1 to (high limit -1) °C
TD High 0.1 to 50°C