Passport 12/Passport 8 Operator’s Manual C-11
C.1.10 AG
Item Name
Configurable
Default
In Config Mode In Monitor Mode
Alarm Yes Yes On
Alm Lev Yes Yes Med
Sweep Yes Yes 6.25 mm/s
O
2
Compen Yes Yes
OR:
General, ICU, NICU, CCU:
On
Off
Operating Mode Yes Yes Measure
Flow Rate Yes Yes
Adu, Ped:
Neo:
120 ml/min
70 ml/min
Auto Standby Yes Yes Off
Apnea Time Yes Yes 20 s
RR Source No Yes Auto
CO
2
Setup
Wave Type Yes Yes Draw
Scale Yes Yes
When unit is mmHg:
When unit is % or KPa:
50
7.0
EtCO
2
High(mmHg)
Yes Yes
Adu, Ped:
Neo:
50
45
EtCO
2
Low(mmHg)
Yes Yes
Adu, Ped:
Neo:
25
30
FiCO
2
High(mmHg)
Yes Yes 4
RR High Yes Yes
Adu, Ped:
Neo:
30
100
RR Low Yes Yes
Adu, Ped:
Neo:
8
30
Gas Setup
Agent Yes Yes AA
N
2
O Scale Yes Yes 50
O
2
Scale Yes Yes
When unit is mmHg:
When unit is % or KPa:
400
50
AA Scale Yes Yes 9.0
Hal/Enf/Iso Scale Yes Yes 2.5
Des Scale Yes Yes 9.0
Sev Scale Yes Yes 4.0
EtO
2
High Yes Yes 88
EtO
2
Low Yes Yes 18
FiO
2
High Yes Yes
Adu, Ped:
Neo:
100
90
FiO
2
Low Yes Yes 18
EtN
2
O High Yes Yes 55
EtN
2
O Low Yes Yes 0
FiN
2
O High Yes Yes 53

Table of Contents

Other manuals for Mindray Passport 12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport 12 and is the answer not in the manual?

Mindray Passport 12 Specifications

General IconGeneral
BrandMindray
ModelPassport 12
CategoryMedical Equipment
LanguageEnglish

Related product manuals