EasyManua.ls Logo

Mindray Passport 12

Mindray Passport 12
284 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Passport 12/Passport 8 Operators Manual C-11
C.1.10 AG
Item Name
Configurable
Default
In Config Mode In Monitor Mode
Alarm Yes Yes On
Alm Lev Yes Yes Med
Sweep Yes Yes 6.25 mm/s
O
2
Compen Yes Yes
OR:
General, ICU, NICU, CCU:
On
Off
Operating Mode Yes Yes Measure
Flow Rate Yes Yes
Adu, Ped:
Neo:
120 ml/min
70 ml/min
Auto Standby Yes Yes Off
Apnea Time Yes Yes 20 s
RR Source No Yes Auto
CO
2
Setup
Wave Type Yes Yes Draw
Scale Yes Yes
When unit is mmHg:
When unit is % or KPa:
50
7.0
EtCO
2
HighmmHg
Yes Yes
Adu, Ped:
Neo:
50
45
EtCO
2
LowmmHg
Yes Yes
Adu, Ped:
Neo:
25
30
FiCO
2
HighmmHg
Yes Yes 4
RR High Yes Yes
Adu, Ped:
Neo:
30
100
RR Low Yes Yes
Adu, Ped:
Neo:
8
30
Gas Setup
Agent Yes Yes AA
N
2
O Scale Yes Yes 50
O
2
Scale Yes Yes
When unit is mmHg:
When unit is % or KPa:
400
50
AA Scale Yes Yes 9.0
Hal/Enf/Iso Scale Yes Yes 2.5
Des Scale Yes Yes 9.0
Sev Scale Yes Yes 4.0
EtO
2
High Yes Yes 88
EtO
2
Low Yes Yes 18
FiO
2
High Yes Yes
Adu, Ped:
Neo:
100
90
FiO
2
Low Yes Yes 18
EtN
2
O High Yes Yes 55
EtN
2
O Low Yes Yes 0
FiN
2
O High Yes Yes 53

Table of Contents

Other manuals for Mindray Passport 12

Related product manuals