Passport 12/Passport 8 Operator’s Manual D-3
D.2 Technical Alarm Messages
Measurement Alarm message L I Cause and solution
XX
XX SelfTest Err H C
An error occurred to the XX module, or there is a
problem with the communications between the
module and the monitor. Re-plug the module and
restart the monitor, or plug the module into another
monitor.
XX Init Err H A
XX Init Err N H A
N is within 1 to 8
XX Comm Err H A
XX Comm Stop H C
XX Limit Err L C
XX parameter limit is accidentally changed. Contact
your service personnel.
XX Overrange L C
The measured XX value is not within the specified
range for XX measurement. Contact your service
personnel.
ECG
ECG Lead Off L* B
The electrode has become detached from the patient
or the lead wire has become disconnected from the
adapter cable. Check the connections of the electrodes
and leadwires.
ECG XX Lead Off L* B
Note: XX represents the leadwires, V, LL, LA,
RA, as per AHA standard or C, F, L and R as
per IEC standard.
ECG Noisy L A
The ECG signal is noisy. Check for any possible sources
of signal noise around the cable and electrode, and
check the patient for great motion.
ECG Artifact (for Mortara ECG
algorithm only)
L A
Artifacts are detected on the ECG analysis lead and as a
result heart rate cannot be calculated and Asystole,
Vfib and Vtac cannot be analyzed. Check the
connections of the electrodes and leadwires and check
for any possible source of interference around the
cable and electrode. Check the patient’s condition and
check the patient for great motion.
ECG High Freq. Noise L A
High frequency signals are detected on the ECG
analysis lead. Check for any possible source of
interference around the cable and electrode.
ECG Low Freq. Noise L A
Low frequency signals are detected on the ECG
analysis lead. Check for any possible source of
interference around the cable and electrode.
ECG Amplitude Too Small L C
The ECG amplitude didn’t reach the detected
threshold. Check for any possible source of
interference around the cable and electrode.
ECG Config. Err L C
ECG configuration is wrongly downloaded. Check the
downloaded configuration and re-download the
correct configuration.

Table of Contents

Other manuals for Mindray Passport 12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport 12 and is the answer not in the manual?

Mindray Passport 12 Specifications

General IconGeneral
BrandMindray
ModelPassport 12
CategoryMedical Equipment
LanguageEnglish

Related product manuals