Recording
8-7
Arrhythmia alarm recording
1. Select the ECG label in the ECG parameter window to the ECG SETUP menu
pops up.
2. Select the ARR ANALYSIS >> option in the ECG SETUP menu, and a popup
menu is opened.
3. Select ARR RECALL>> to open the ARR RECALL window.
4. Select the WAVE >> option to open the ARR WAVE RECALL window
5. Select REC to start recording.
6. When the recording completes, the recorder stops automatically.
CO measurement curve recording
1. Select the CO label in the parameter area to open the C.O. SELECT menu.
2. Select C.O. MEASURE option to open the WINDOW FOR CO
MEASUREMENT.
3. Press START option to start a CO measurement.
4. Press REC option to start recording.
5. When the recording completes, the recorder stops automatically.
Hemodynamic calculation recording
1. Select EDIT>> in the WINDOW FOR CO MEASUREMENT to open the
WINDOW FOR C.O. EDIT.
2. Select HEMO CALCULATE>> to open the HEMOD WINDOW.
3. Input the value of corresponding parameters, and then select CALCULATE
option to start the calculation.
4. Press REC option to start recording.
5. When the recording completes, the recorder stops automatically.
Titration table recording
1. Select DRUG CALC >> in SYSTEM MENU.
2. Perform the drug calculation, and then select the TITRATION >> option.
3. Select the REC option to start recording.
4. When the recording completes, the recorder stops automatically.