Appendices
20-15
A.13 IBP Specifications
Number of channels 2
Pressure labels ART, PA, CVP, RAP, LAP, ICP, P1 and P2
ART 0 to 300 mmHg
PA – 6 to 120 mmHg
CVP/RAP/LAP/ICP – 10 to 40 mmHg
Measurement range
P1/P2 – 50 to 300 mmHg
Resolution 1 mmHg
Precision ±2% or ±1mmHg, whichever is greater
Update period 1s
Pressure transducer
Sensitivity 5 uV/V/mmHg
Impedance range 300 to 3000