ECG/RESP Monitoring
11-26
11.5.2 Arrhythmia Analysis Menu
Select ARR ANALYSIS >> in ECG SETUP menu. The following menu appears.
Figure 11-13 Arrhythmia Analysis
In this menu, you can perform the following settings:
ARR ANAL
Arrhythmia analysis
ON: Enables the arrhythmia analysis;
OFF: Disables the arrhythmia analysis.
PVCs ALM
PVCs alarm
ON: When the PVCs alarm occurs, the monitor gives alarm
indications and saves the alarm;
OFF: When the PVCs alarm occurs, the monitor neither gives
alarm indications nor saves the alarm;
When OFF is selected, the
icon is displayed on the right of
PVCs in the parameter window.
ALM LEV
Alarm level
Options: HIGH, MED and LOW.
ALM REC
Alarm recording
ON: The monitor starts recording when the PVCs alarm occurs.
OFF: The monitor does not record when the PVCs alarm
occurs.
ALM HI
Upper alarm limit
Determines the upper limit of the PVCs alarm. Range: 110.
An alarm is triggered when the PVCs exceeds the upper limit.
ARR
RELEARN
Arrhythmia relearning
You can select this option to start a learning procedure. The
“ARR LEARNING” message is displayed in the information
area of the screen.