20-2
Appendix A Product Specifications
A.1 Safety Classifications
Type of protection against
electric shock
Class I with internal electric power supply.
Where the integrity of the external protective earth (ground) in
the installation or its conductors is in doubt, the equipment shall
be operated from its internal electric power supply (batteries)
Degree of protection
against electric shock
Sidestream CO
2
/AG:
ECG/RESP/TEMP/SpO
2
/NIBP/IBP/
CO/mainstream CO
2
:
BF (defibrillation proof)
CF (defibrillation proof)
Degree of protection
against hazards of ignition
of flammable anesthetic
mixtures
Not protected (ordinary)
Degree of protection
against harmful ingress of
water
Not protected (ordinary)
Mode of operation Continuous
Equipment type Portable

Table of Contents