EasyManua.ls Logo

Mindray PM-9000

Mindray PM-9000
378 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendices
20-14
A.12 TEMP Specifications
Number of channels 2
Displayed parameters T1, T2 and TD
Measurement range
0 to 50°C (32 to 122°F)
Resolution
0.1°C
Precision
0.1°C (excluding the sensor)
±0.2°C (including the YSI 400 series sensor)
Update period 1s
Minimum time for
accurate measurement
Body surface: < 100s
Body cavity: < 80s
(YSI 400 series sensor)

Table of Contents

Related product manuals