Appendices
20-14
A.12 TEMP Specifications
Number of channels 2
Displayed parameters T1, T2 and TD
Measurement range
0 to 50°C (32 to 122°F)
Resolution
0.1°C
Precision
0.1°C (excluding the sensor)
±0.2°C (including the YSI 400 series sensor)
Update period 1s
Minimum time for
accurate measurement
Body surface: < 100s
Body cavity: < 80s
(YSI 400 series sensor)