EasyManua.ls Logo

Mindray PM-9000 - Precautions

Mindray PM-9000
378 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SpO
2
Monitoring
12-14
%SpO
2
100
97
95
90
85
80
75
70
65
DST
TM
Relative Correlation Canceler
Energy Output
Discrete Saturation Transform
The DST plot has two peaks: the peak corresponding to the higher saturation is
selected as the SpO
2
value. This entire sequence is repeated once every two seconds
on the most recent four seconds of raw data. The SpO
2
value therefore corresponds
to a running average of arterial hemoglobin saturation that is updated every two
seconds.
12.3.2 Precautions
WARNING
z The pulse wave from the MS-7 module must NOT be used for apnea
monitoring.
z As a trend towards patient deoxygenation is indicated, blood samples
should be analyzed by a laboratory co-oximeter to completely
understand the patient’s condition.
z If an alarm condition (other than exceptions listed herein) occurs in the
Alarms Silenced status, the monitor only gives visual alarm symbols.
z Measure the monitor’s leakage current whenever an external device is
connected to the serial port. Leakage current must not exceed 100
microamperes.

Table of Contents

Related product manuals