Installation and Maintenance
3-11
3.2.3 Disinfection
Disinfection may cause damage to the equipment. We recommend the disinfection is
contained in the hospital’s servicing schedule only when necessary. The equipment
should be cleaned prior to disinfection.
Recommended disinfection material: Alcohol based (Ethanol 70%, Isopropanol
70%), and aldehyde based.
WARNING
z Disinfection may cause damage to the equipment; therefore, when
preparing to disinfect the equipment, consult your hospital’s infection
controllers or professionals.
z The cleaning solutions above can only be used for general cleaning. If
you use them to control infections, we shall assume no responsible for
the effectiveness.
NOTE
z ALWAYS dilute the solutions according to the manufacturer’s
suggestions and adopt lower concentration if possible.
z NEVER submerge the equipment into water or any solution, or pour
water or any solution on the equipment;
z ALWAYS wipe off all the excess liquids on the equipment surface and
accessory surface with a dry cloth;
z Never use EtO and formaldehyde to disinfect.
z Never permit high-pressure and high-temperature disinfection of the
equipment and accessories.