SpO
2
Monitoring
12-10
ALM
SpO
2
alarm on/off status
ON: When a SpO
2
alarm occurs, the monitor gives alarm
indications and stores the alarm;
OFF: When a SpO
2
alarm occurs, the monitor neither gives
alarm indications nor stores the alarm;
When OFF is selected, the
icon is displayed on the right
of the SPO2 label.
ALM LEV
Alarm level
Options: HIGH, MED and LOW.
ALM REC
Alarm recording
ON: When a SpO
2
alarm occurs, the monitor enables the
recording;
OFF: When a SpO
2
alarm occurs, the monitor does not
enable the recording.
SPO2 ALM HI
SpO
2
upper alarm limit
SPO2 ALM LO
SpO
2
lower alarm limit
PR ALM HI
PR upper alarm limit
PR ALM LO
PR lower alarm limit
SpO
2
and PR alarm limits:
Parameter Max. upper limit Min. lower limit Step
SpO
2
100 0 1
PR 254 0 1
The default SpO
2
and PR alarm limits:
Parameter Patient type Upper limit Lower limit
Adult 100 90
Pediatric 100 90
SpO
2
Neonate 95 80
Adult 120 50
Pediatric 160 75
PR
Neonate 200 100