NIBP Monitoring
13-8
îš„ SYS ALM LO
Determines the lower limit of the systolic pressure.
îš„ MEAN ALM HI
Determines the upper limit of the mean pressure.
îš„ MEAN ALM LO
Determines the lower limit of the mean pressure.
îš„ DIA ALM HI
Determines the upper limit of the diastolic pressure.
îš„ DIA ALM LO
Determines the lower limit of the diastolic pressure.
If a measured pressure crosses a preset upper or lower alarm limit, an alarm will be
triggered. The NIBP alarm limits are as follows:
Patient type Adult Pediatric Neonate
Systolic pressure 40–270 mmHg 40–200 mmHg 40–135 mmHg
Mean pressure 20–235 mmHg 20–165 mmHg 20–110 mmHg
Diastolic pressure 10–215 mmHg 10–150 mmHg 10–100 mmHg
îš„ DISPLAY WAY
1 GROUP: The NIBP parameter area only displays a group
of NIBP values obtained in the latest measurement.
GROUPS: Multiple groups of measurement values are
displayed at the lower left corner of the screen as shown
below.
NS NM ND PR DATE TIME
108 84 70 68 2006-6-12 16:06:00
108 80 70 67 2006-6-12 16:11:00
108 80 70 72 2006-6-12 16:16:00
îš„ UNIT
Options: mmHg, kPa;
îš„ INTERVAL
Select MANUAL to set the monitor to manual NIBP
measurement mode, or select from the time options to
determine the interval between automatic measurements.
Optional intervals: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180,
240, and 480MIN.
îš„ RESET
Select this option to restore the initial settings of the
pressure pump. If the monitor fails to give a visual
indication when the pressure pump is working improperly,
selecting this option activates a self-test procedure, and
restores the monitor to normal performance.
îš„ CONTINUAL
Select this option to start a continuous measurement for five
minutes.