CO Monitoring
16-10
îš„ ALM REC
Alarm recording
ON: When a TB alarm occurs, the monitor enables the
recording;
OFF: When a TB alarm occurs, the monitor does not enable
the recording.
îš„ TB ALM HI
TB upper alarm limit
îš„ TB ALM LO
TB lower alarm limit
TB alarm limits:
Parameter Maxi. Upper Mini. Lower Step
TB 43 23 0.1
îš„ CO.CONST
This coefficient is related to the catheter and the injectate
volume. After replacing the catheter, you should adjust this
coefficient according to the instructions for use.
îš„ INJ. TEMP
FROM
Injectate temperature source
ON: When this option is selected, the system obtains the
real-time injectate temperature by placing a sensor in the
way of the injectate.
OFF: When this option is selected, the injectate temperature
is given and you can set INJ. TEMP in the menu. The
system obtains the preset injectate temperature directly from
the setting, instead of real-time detection.
îš„ INJ. TEMP
Injectate temperature
When the INJ. TEMP FROM option is set to OFF, you can
adjust this option to set the injectate temperature.
Range : 0 to 27℃.
îš„ TEMP UNIT
Temperature unit
Options: ℃ and ℉.
îš„ INT TIME(s)
The minimum interval between two consecutive
measurements
Unit: second.
Range: 5 to 300s.
The time increases or decreases by 5 seconds if you rotate
the control knob for one time.
îš„ DEFAULT >>
You can select this option to access the CO DEFAULT
CONFIG menu, in which you may select FACTORY
DEFAULT CONFIG or USER DEFAULT CONFIG. After
finishing your selection and exiting the menu, the system
pops up a dialog box asking for your confirmation.