CO
2
Monitoring
17-8
Other Setup
Selecting OTHER SETUP option opens the following menu.
Figure 17-4 CO
2
Other Setup Menu
You can perform the following settings in this menu.
îš„ PUMP RATE
Determines the sampling rate of the CO
2
module pump.
Options: 150ml/min, 100ml/min
îš„ N2O COMPEN
N
2
O compensation
ON: Select ON, when the N
2
O content in the measured gas
is no smaller than 30%.
OFF: Select OFF, when the N
2
O content in the measured
gas is smaller than 30%.
îš„ O2 COMPEN
O
2
compensation
ON: Select ON, when the O
2
content in the measured gas is
no smaller than 50%.
OFF: Select OFF, when the O
2
content in the measured gas
is smaller than 50%.
îš„ Des COMPEN
Desflurane compensation
ON: Select ON, when the desflurane is contained in the
measured gas.
OFF: Select OFF, when the deflurane is not contained in the
measured gas.
îš„ BTPS
Body Temperature and Pressure, Saturated
Options: ON and OFF.