CO
2
Monitoring
17-18
NOTE
z When measuring a moist gas of saturated water vapor at the body
temperature and the ambient pressure, turn ON BTPS; when measuring
a dry gas at the ambient temperature and pressure, turn OFF BTPS.
îš„ AUTO
STANDBY
Range: 0 to 60min.
If no respiration waveform is detected in the selected time,
the CO
2
module enters the standby mode automatically.
When AUTO STANDBY is set to 0min, it means the CO
2
module is in MEASURE state. If the sampling line is not
connected, the CO
2
module enters the standby mode
automatically 3 minutes later.
îš„ ZERO CAL
This option allows you to zero manually so as to eliminate
the adverse affect of baseline drift in the process of
measurement.
îš„ DEFAULT >>
Select DEFAULT >> to access the CO2 DEFAULT
CONFIG menu. You can select either FACTORY
DEFAULT CONF or USER DEFAULT CONF. After
finishing the selection and exiting the menu, a dialog pops
up asking for confirmation of your selection.