CO
2
Monitoring
17-25
îš„ ALM LEV
Alarm level
Options: HIGH, MED and LOW.
îš„ ALM REC
Alarm recording
ON: When a CO
2
alarm occurs, the monitor enables the
recording;
OFF: When a CO
2
alarm occurs, the monitor does not
enable the recording
îš„ CO2 ALM HI
Upper alarm limit of EtCO
2
îš„ CO2 ALM LO
Lower alarm limit of EtCO
2
îš„ INS ALM HI
Upper alarm limit of InsCO
2
îš„ AWRR ALM HI
Upper alarm limit of AwRR
îš„ AWRR ALM LO
Lower alarm limit of AwRR
îš„ APNEA ALM
Determines the apnea alarm delay.
If the apnea of the patient exceeds the preset apnea alarm
delay, the monitor triggers an alarm and gives the " CO2
APNEA " alarm message.
Options: 10S, 15S, 20S, 25S, 30S, 35S and 40S.
îš„ SWEEP
Waveform speed
Options: 6.25, 12.5 and 25.0.
Unit: mm/s.
îš„ UNIT
Options: mmHg and kPa.
îš„ WAVE SCALE
Options: HIGH and LOW.
This option allows you to adjust the amplitude of the CO
2
waveform.
îš„ WORK MODE
Options: MEASURE and STANDBY.
To start the CO
2
monitoring, select the MEASURE mode.
In the STANDBY mode, the air pump and infrared ray
source of the CO
2
module are disabled. Hence, decreases
the power consumption and prolongs the operating life of
infrared ray source and the whole CO
2
module.
NOTE
z The apnea alarm cannot be disabled.
z When the CO
2
monitoring is not required, it is recommended the sensor
not be connected and the work mode be set to STANDBY.