Anesthesia Gas Monitoring
18-7
SWEEP
Used to select the speed to scan the screen waveforms.
Options: 6.25mm/s and 12.5mm/s.
WORK MODE
Options: MEASURE and STANDBY.
To monitor the anesthetic gas, select the MEASURE option. Otherwise, select the
STANDBY option.
Alarm Setup Menu
Select ALARM SETUP >> in AG SETUP menu. The following menu appears.
Figure 18-5 Alarm Setup Menu
In the ALARM SETUP menu-1, you can set the following items.
ALM
Alarm on/off status
ON: When the AG has an alarm, the system will give alarm
indications and save the alarm information.
OFF: The system will not give alarm indications when an
AG alarm occurs. Instead it will display
on the right of
the GAS label in the parameter window.
ALM LEV
Options: HIGH, MED and LOW.
ALM REC
ON: The recorder outputs the alarm information in case an
AG alarm occurs.
OFF: The recorder does not output the alarm information in
case an AG alarm occurs.
EtCO2 ALM HI
Set the upper limit of EtCO
2
which triggers the alarm