Appendices
20-12
Resolution 1bpm
Precision
±3bpm (non-motion conditions)
±5bpm (in motion conditions)
Refreshing rate 1s
A.10.3 Nellcor SpO
2
Specifications
Sensor Range Precision*
MAX-A, MAX-AL, MAX-N,
MAX-P, MAX-I and MAX-FAST
70 to 100%
0% to 69%
±2%
Undefined
OxiCliq A, OxiCliq N, OxiCliq P,
OxiCliq I
70 to 100%
0% to 69%
±2.5%
Undefined
D-YS, DS-100A, OXI-A/N and
OXI-P/I
70 to 100%
0% to 69%
±3%
Undefined
SpO
2
measurement range
and precision
MAX-R, D-YSE and D-YSPD
70 to 100%
0% to 69%
±3.5%
Undefined
PR measurement range and
precision
20 to 250bpm: ±3bpm
251 to 300bpm: Undefined
Refreshing rate 1s
*: When sensors are used on neonatal subjects as recommended, the specified precision range is
increased by ±1%, to account for the theoretical effect on oximeter measurements of fetal
hemoglobin in neonatal blood.