EasyManua.ls Logo

Mindray PM-9000 - Page 347

Mindray PM-9000
378 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendices
20-21
15 to 18
±1
>18 Not specified
0 to 1
±0.15
1 to 5
±0.2
5 to 8
±0.4
Sev
> 8 Not specified
0 to 1
±0.15
1 to 5
±0.2
Enf, Iso, Hal
> 5 Not specified
0 to 25
±1
25 to 80
±2
O2 (Optional)
80 to 100
±3
2 to 60 BrPM
±1 BrPM
AwRR
> 60 BrPM Not specified
Alarm range
CO2:
AwRR:
0 to 10 % (0 - 76 mmHg)
2 to 100 BrPM
Apnea alarm delay AwRR 20 - 40 s
Refreshing rate 1s
Calibration Yearly calibration requested.
Calibration stability
After module being used for 12 consective months, the error is <
1%
CO
2
250 ms (fall time 200 ms)
N
2
O 250 ms
O
2
600ms
HAL, ISO, SEV, DES 300 ms
Rise time (10 % to 90 %)
Sampling flow 120ml/min,
using the DRYLINE™
water trap and neonatal
DRYLINE™ sampling
line (2.5m)
ENF 350 ms
CO
2
250 ms (fall time 200 ms)
N
2
O 250 ms
O
2
500ms
HAL, ISO, SEV, DES 300 ms
Rise time (10 % to 90 %)
Sampling flow 200ml/min,
using the DRYLINE™
water trap and adult
DRYLINE™ sampling
line (2.5m)
ENF 350 ms
Delay time < 4s

Table of Contents

Related product manuals