System Menu
4-12
îš„ ALM SEL
Alarm selection
Options: COMMON ALM SETUP, XX ALM SETUP;
(XX refers to HR, ST, PVCs, SPO2, NIBP, IBP (1,2),
CO2, RESP, TEMP, CO and AG).
îš„ ALARM VOL
Alarm volume
The volume can be set between 0 and 10. 0 means off
and 10 is the maximum volume.
îš„ ALM REC TIME
Alarm recording time
Options: 8S,16S and 32S.
When an alarm occurs, the recorder records according
to the alarm recording time.
îš„ ALM PAUSED TIME Options: 1MIN, 2MIN and 3MIN.
îš„ PARA ALAM TYPE Options : LATCH and UNLATCH.
Detailed information about alarms is given in chapter 6 Alarms.
If a parameter alarm setup is selected from the drop-down list of ALM SEL, the
corresponding alarm setup items will be displayed in the ALARM SETUP menu.
When the ALARM VOL is set to 0, data will not be saved upon power failure. After
the monitor restarts, the ALARM VOL restores the default.
When the ALARM VOL is set to 0, it will restore its default if you silence or pause
alarms.