System Menu
4-30
îš„ ALM LEV
Alarm level
Options: HIGH, MED (medium) and LOW.
More than one option can be selected at one time.
îš„ ALM TYPE
Alarm type
Options: TECH. (technical) and PHYS. (physiological).
Both options can be selected at one time.
Trigger Conditions
A nurse call signal will be triggered only if all the following conditions are met:
1. The nurse call function is enabled.
2. An alarm of the preset alarm level and alarm type comes out.
3. The monitor is not in the Alarms Paused or the System Silenced status.
NOTE
z If no option in ALM LEV or ALM TYPE is selected, the nurse call signal
will not be triggered in whatever condition.
z The nurse call function can’t be used as the main alarm notice method.
Medical staff must combine the audible and visual alarms, the clinical
vital signs of the patient to determine the patient’s situation.
z In the Alarms Paused or the System Silenced status, the nurse call
function of the monitor will be disabled automatically.