4 Setup
Operator’s Manual 4 - 45
If the patient data are already stored in the hard disk, the system will pop-up prompts, follow
the steps below:
a. Click [OK].
b. Click [Wipe] or [Secure Wipe] and operate according to the screen prompts to clear
patient data.
If [Secure Wipe] is selected, the deleted patient data cannot be restored with restoration
tools. Perform this operation with caution.
c. Select [User-defined] again, and click [Confirm].
d. Input the password and click [Confirm] to finish the password setting.
NOTE:
If you want to switch to Factory Default, perform steps above again. The password is the same
as that of the User Define.
When you set password, multi-language and Chinese characters are not supported.
4.11.2 Anti-Virus
The system provides two pieces of anti-virus software: Windows Defender and McAfee. They can
effectively prevent the ultrasound system from being attacked by virus, spyware, or other malware.
Windows Defender is already installed on the system.
The McAfee software is an option. If the McAfee software is installed, the system displays
“McAfee is running”. If you want to buy McAfee, contact Mindray representatives.
NOTE:
McAfee cannot be uninstalled after successful installation.
After the option is installed, the system will be restarted. And after restart, the system installs
McAfee automatically, which needs about one hour.
If McAfee is improperly installed due to power-off, shut-down, closing of cmd.exe, or any
other abnormal operation during the installation, please contact the Mindray service engineers.
4.12 System Information
Click [About] on the Setup menu to enter the system information screen.
About
The “About” shows the information about the product type, manufacturer, configuration type,
ECN/TCN, version number, MAC address, etc.
You cannot edit the information, but only view them. The information varies depending on the
system configurations and version.
Statistics
Log onto the system as Administrator, and click [Statistics] to enter the statistics screen, which
shows the total exam numbers of each probe, application and operator.
Open Source
The “Open Source License” and “Open Source Info.” pages display the license and declaration
of open sources.

Table of Contents

Other manuals for Mindray Resona 7T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Resona 7T and is the answer not in the manual?

Mindray Resona 7T Specifications

General IconGeneral
BrandMindray
ModelResona 7T
CategoryMedical Equipment
LanguageEnglish

Related product manuals