EasyManuals Logo

Mindray Resona 7T User Manual

Mindray Resona 7T
492 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #40 background imageLoading...
Page #40 background image
2 - 12 Operator’s Manual
2 System Overview
NOTE:
S-VIDEO port performs better in analog video print.
2.9 Power Supply Panel
Figure 2-3 Power Supply Panel
9. / 5 V indicator
10. / 3.3 V indicator
11. / PC_LVDS indicator
12. / Preserved function, PHV_OP protection indicator.
13. Battery status indicator:
• Charging status:
It illuminates in green when batteries are charged fully.
• Discharging status:
It illuminates in green when the power of the batteries is
more than 20%;
It blinks in orange when the power of the batteries is lower
than 20%.
It blinks quickly in orange when the power of the batteries
is lower than 5%.
No. Name Function
No. Name Function
1. Alternative current
auxiliary output
Supply power for optional peripheral devices.
2. Circuit breaker Used for switching off/ on the power supply.
3. Power inlet AC power inlet
4. Equipotential
terminal
Used for equipotential connection, that balances the
protective earth potentials between the system and other
electrical equipment.

Table of Contents

Other manuals for Mindray Resona 7T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Resona 7T and is the answer not in the manual?

Mindray Resona 7T Specifications

General IconGeneral
BrandMindray
ModelResona 7T
CategoryMedical Equipment
LanguageEnglish

Related product manuals