10 - 4 Operator’s Manual
10 Physiological Unit Signal
a. Tap [RESP].
b. Adjust [Speed], [RESP Gain], [Position] and [Invert].
5. Exit Respiratory display mode, and remove ECG electrodes from the patient.
6. Tap [Physio] or press the user-defined <Physio> key to exit physio mode.
10.3 PCG
NOTE:
• Check PCG each time before use.If there are any cracks or scratch, please contact the sales
representative.
• If there is no ECG signal (PCG waveform is flat) after entering ECG mode, please check
whether the sensor plug is connected well.
• Do not hot unplug the PCG sensor.
• To satisfy the PCG sensor's receiving performance, do not rub the acoustic window hard
because of the vulnerability of the silicone.
10.3.1 PCG Operation Basic Procedures
Perform the following procedure:
1. Connect PCG sensor.
a. Connect the PCG sensor to the corresponding interface on the physiological module.
b. Power on the system.
c. Place PCG transducer on the patient.
The position is shown below:
Auscultation area of the mitral valve: the fifth intercostal space on the left, the interior of
clavicle’s middle line.
Auscultation area of tricuspid valve: the right of the sternum.
Auscultation area of aortic valve: the second intercostal space of the right sternum.
Auscultation area of pulmonary valve: the second intercostal space of the left sternum.
1 Aortic valve auscultation area 2 Tricuspid valve auscultation area
1
2
3
4

Table of Contents

Other manuals for Mindray Resona 7T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Resona 7T and is the answer not in the manual?

Mindray Resona 7T Specifications

General IconGeneral
BrandMindray
ModelResona 7T
CategoryMedical Equipment
LanguageEnglish

Related product manuals