9-32 Structure and Assembly/Disassembly
8. Unscrew 4 M4 X12 cross panhead screws with screwdriver (M3, M4) to remove the connection
wires of the coupling gel heating cup.
9. Unscrew 2 M5 X 16 round inner head screw (with the pad) from the stop block by M5 spanner.
10. Unscrew six M5 X 16 round inner head screws from the keyboard base with the M6 inner
hexagon spanner. Rotate the base assembly anticlockwise until exposing the electrical magnet
(electromagnet 120 kgf) (separating the electromagnet), and the lift base assembly straight
(FRU).
9.3.15 Control Panel Moving Mechanism Assembly
Note: the ultrasound system can be disassembled only for the system equipped with the moving
mechanism.
The disassembly tool: cross-headed screwdriver (M3, M4), flat-headed screwdriver, inner hex
spanner (M3, M5).
1. See Chapter 9.3.6, 9.3.8 and 9.3.9 for details.
2. Unscrew the screws in the following figures, and unplug the cables. See Chapter 9.3.13 for
details.
B
ase assembly
(FRU)
Electrical magnet