3 - 4 Operator’s Manual
3 System Preparation
You must log in again after system restart or dormancy.
To Login onto the System
Perform the following procedure:
1. Select the login type (Local or LDAP), and user name in the drop-down list.
2. Enter the password and click [Login].
When the user has logged onto the system, is visible in the bottom-right corner of the
screen.
To Change Users
Perform the following procedure:
1. Click in the bottom-right corner of the screen.
2. Click [Change User] to bring up the Login dialog box.
3. Select the login type (Local or LDAP), and user name in the drop-down list.
4. Enter the password and click [Login].
To Modify Password
General operators and administrators can modify the password.
Perform the following procedure:
1. Click in the bottom-right corner to bring up the Session Manage dialog box where you can
see the current user’s information.
2. If you want to modify the current password, click [Change Password].
3. Enter both the previous and new passwords, and confirm the new password in the dialog box.
4. Click [OK] to exit.
To Lock the System
Perform the following procedure:
1. Click the in the bottom-right corner of the screen to bring up the dialog box.
2. Select [Lock Machine] and the system is locked.
You must log on before using the system.
3.3.3 Checking After Powering On
Check after the system is powered on:
• There are no unusual sounds or smells indicating possible overheating.
• There are no persistently displayed system error messages.
• There is no evident excessive noise, or discontinuous, absent or black items in the B mode
image.
• Check whether there is abnormal heat on the surface of the probe during an ultrasound
procedure.
• The control panel keys and knobs are fully functional.
• The date and time are displayed correctly.
• The touch screen and the main monitor screens display normally depending on the system
modes and image status.