Calibration Calibration Procedure
4 - 18 0070-10-0556-02 Spectrum
®
/Spectrum OR
Service Manual
NOTE: If no gas is being delivered, or the mixture does not contain
5% CO
2
, the message “Calibration error. Caused by no gas
or wrong gas concentration” will appear. Obtain a new gas
canister and return step 1.
7. When the proper gas mixture is applied, the message Calibrating, continue to
apply 5% CO
2
will appear in the Calibration Menu window. When the calibration
is complete, the message will change to Calculating, calibration gas can be
removed. Release the button, on the canister, and remove the connector.
8. After a moment, the message will change to Calibration Completed Successfully.
The date and time of the successful calibration will appear in the Calibration Menu.
9. Use the Navigator
Knob to select Previous Menu.
10. Use the Navigator Knob to select Normal Screen and return to the monitor’s normal
display screen.
11. When the Spectrum
®
/Spectrum OR
has detected valid breaths, data will display
for the CO
2
, Inspired CO
2
and Respiratory Rate.
12. The CO
2
respiration waveform and data will automatically replace the ECG respiration
waveform and data on the display. If respiration wave or data is not displayed, use the
Display Setup Menu to select RESP or CO
2
to be displayed as desired.
13. The CO
2
waveform scale can be changed by accessing the CO
2
Menu.
NOTE: Microstream
®
CO
2
waste and CO
2
FilterLine
®
should be
treated as biohazardous waste.