Service Manual 0070-10-0556-02 4 - 19
Calibration Procedure Verification
4.6 Verification
4.6.1 Initial Set-up
1. Using a patient simulator, connect the ECG, IBP1, IBP2 and temperature cables to the
left side connector panel. Set the ECG simulator for 60 bpms, 1mv QRS signal.
2. Set up the Patient Menu for Adult (Patient Size) as follows:
A. Monitor Setup 1. Display Setup – 3 Waveforms
2. ECG Speed – 25 mm/sec
3. IBP Speed – 25 mm/sec (optional)
4. Respiration / Gas Speed – 12.5 mm/sec
5. EEG Speed – 25 mm/sec
B. Print Setup 1. Waveform 1 – ECG 1
2. Waveform 2 – ECG 2
3. Select Printer – Local
C. Parameters 1. ECG:
a. ECG 1 – II
b. ECG 2 – I
c. ECG 3 – III
d. ECG 1 thru 6 Size - 1 cm/mV (12 Lead) Page 1
e. ECG 1 - I
f. ECG 2 - II
g. ECG 3 - III
h. ECG 4 - AVR
i. ECG 5 - AVL
j. ECG 6 - AVF (12 Lead) Page 2
k. ECG 1 - V1
l. ECG 2 - V2
m. ECG 3 - V3
n. ECG 4 - V4
o. ECG 5 - V5
p. ECG 6 - V6
NOTE: 12 Lead applies to Spectrum only.