Verification Calibration Procedure
4 - 20 0070-10-0556-02 Spectrum
®
/Spectrum OR
™
Service Manual
D. NIBP 1. Set Start Pressure – 180 mmHg
Interval – 5 minutes
2. IBP1 – Scale 0 to 160 mmHg
3. IBP2 – Scale 0 to 80 mmHg
4. SpO
2
a. Averaging mode – 2
b. Sensor Off Audio – off
5. CO
2
(optional)
a. Apnea Delay – 60
b. Scale 40 mmHg
6. Respiration
a. Resp lead – II
b. Apnea Delay – 60
c. Resp source – Auto
d. Scale – 3
7. Gases (optional)
a. Select agent – Auto
b. O
2
scale - 100%
c. N
2
0 Scale – 10%
8. Spirometry Verification (optional)
a. Select display loops - pressure volume
b. Paw Scale - 0-40
c. Volume scale - 0- 900
d. Flow Scale - -40 -+40
9. BISx Verification (optional)
a. Expand view - no
b. EEG setup menu - -25uV to +25uV
c. Sensor Check
•Pass or Fail
• High impedance
• Noise
•Lead off