Spectrum
®
/Spectrum OR
™
Service Manual 0070-10-0556-02 4 - 23
Calibration Procedure Verification
4.6.5 SpO
2
Verification
1. Set the display waveform 4 as Pleth. Set the HR source to Auto.
2. Verify that the SpO
2
message is displaying SpO
2
No Sensor in Message area “B”.
3. Connect the SpO
2
sensor to the panel connector. Verify the SpO
2
message changes to
Sensor Off or SpO
2
Initializing.
4. Apply sensor to finger.
5. Verify SpO
2
displays the pleth waveform, and the SpO
2
indicates a valid reading.
Verify the HR source is SpO
2
and a beep tone is present.
4.6.6 NIBP Verification
1. Connect the Adult cuff connector to the NIBP hose. Attach the NIBP hose to the Cuff
connection the left side on the monitor.
2. Apply cuff and press the START key.
3. Verify the pump motor starts to pump and inflates the cuff to 180 mmhg (Adult).
The cuff will begin to deflate and obtain a blood pressure reading of Sys/ Dia/ Mean in
about 20 to 30 seconds after peak pressure is obtained.
4. Verify the reading on screen.
4.6.7 Battery Operation Verification
1. If batteries are installed in the unit remove them.
2. Verify the unit functions on Line power correctly.
3. Install the two batteries in the appropriate slots located on the left side of the monitor.
4. Remove the line cord from the unit. Verify the unit operation is not interrupted.
5. Remove one of the batteries and verify the unit still operates. Verify the second battery
operates, if installed alone.
4.6.8 CO
2
Operation Verification
1. Connect the FilterLine
®
Short Term assembly to the input port of the CO
2
connector on
the left side of the monitor.
2. Attach a can of Calibration Gas (P/N 0075-00-0033-01) to the Filterline Short Term
assembly. Feed gas into monitor and verify the ETCO
2
, Inspired CO
2
and respiration
readings are displayed on the screen.
4.6.9 Cardiac Output Verification
Setup
1. Access the Installation Menu by pressing and holding the DISCHARGE key
(Spectrum
®
) or the TRENDS key (Spectrum OR
™
) while powering ON the monitor.
2. Select Enable EPM and change to Yes.
3. Select the weight units to kgs.
4. Select the height units to cms.
5. Save the current settings and turn unit OFF.