Spectrum
®
/Spectrum OR
Service Manual 0070-10-0556-02 4 - 1
4.0
Calibration Procedure
4.1 Introduction
The following procedures are provided to verify the proper operation of the Spectrum
®
/
Spectrum OR
Monitor. Service Diagnostics provide the capability of diagnosing
problems within the Spectrum
®
/Spectrum OR
hardware. A menu driven interface
similar to that of the Spectrum
®
/Spectrum OR
User Interface, is used to execute all
tests.
CAUTION: Calibration is not to be performed while monitoring a
patient.
4.2 Warning and Guidelines
In the event that the instrument covers are removed, observe the following warnings and
general guidelines:
1. Do not short component leads together.
2. Perform all steps in the exact order given.
3. Use extreme care when reaching inside the opened instrument. Do not contact exposed
metal parts which may become electrically active.
4. Read and understand each step of the procedure prior to beginning the step.