Care and Cleaning of View 12
ECG Analysis Module Preventative Maintenance
5 - 6 0070-10-0556-02 Spectrum
®
/Spectrum OR
Service Manual
CAUTION: To avoid permanent damage, do not expose metal
components (pins, sockets, snaps) to disinfectants, soaps or
chemicals.
NOTE: ECG cables and leadwires must never be immersed, soaked
in any fluids, and they should not be cleaned with harsh
chemicals such as acetone or non-diluted bleach.
NOTE: Do not autoclave, radiation or steam sterilize ECG cables or
leadwires.
NOTE: Extended exposure to Ethylene Oxide gas may shorten life
of the ECG cables and leadwires, leading to poor signal
quality.
5.10 Care and Cleaning of View 12
ECG Analysis
Module
Clean cables and leadwires using a cloth wipe and warm water. Use a dry, clean cloth to
dry leadwires and cables before placing them on a patient. Do not use alcohol to clean the
View 12 ECG Analysis Module. Alcohol or other harsh chemicals will cause the cables and
leadwires to become brittle or harden, causing damage.
NOTE: The View 12 ECG Analysis Module must never be immersed
or soaked in any fluids.
5.11 Battery Replacement and Maintenance
5.11.1 Battery Replacement
1. Open battery compartment door, on left side of unit, by pressing the finger grip area
and sliding the door to the left.
2. Press the release button, located on the right side of the installed battery. This will eject
the battery. Slide out old battery.
3. Slide in replacement battery until it clicks into place.
4. Close battery compartment door by sliding the door to the right until it firmly clicks into
place.
CAUTION: Replace sealed lead acid batteries with P/N 0146-00-0043
ONLY. Replace lithium-ion batteries with P/N 0146-00-0069
ONLY.
5.11.2 Battery Maintenance
The batteries may be subject to local regulations regarding disposal. At the end of the
battery life, dispose of the batteries in accordance with any local regulations.
CAUTION: Recharge batteries in the Spectrum
®
/Spectrum OR
.
CAUTION: Remove the batteries if the Spectrum
®
/Spectrum OR
is not
likely to be used for an extended period of time.