Spectrum
®
/Spectrum OR
Service Manual 0070-10-0556-02 1 - 25
Repair Information Troubleshooting Menus
NIBP: Retry A measurement has been
attempted but no reading was
possible and the retry limit has not
been reached.
Retry will be attempted.
Check for leaks and quality of
peripheral pulses. Decrease
patient movement. Switch cuff
to another limb.
Unable To Measure An unsuccessful measurement
cycle has been completed.
Switch cuff to another limb.
Decrease patient movement.
Press START to retry. Be
prepared to auscultate BP
manually.
Contact Customer Support.
NIBP: Cuff Overpressure The hardware overpressure limit
has been exceeded.
Power cycle unit.
If message reappears, contact
Customer Support.
NIBP: Cuff Overpressure/
Press STOP to clear.
The hardware overpressure limit
has been exceeded.
Press STOP to clear the
hardware overpressure.
If message reappears, contact
Customer Support.
NIBP: Check Calibration The software has detected that the
overpressure transducer is out of
calibration.
Have the unit calibrated.
If problem persists contact
Customer Support.
Unable to obtain a BP Patient movement Wait until patient is calm or
gently hold limb.
Cuff or hose NOT attached /
leaking
Check all connections.
HR irregular / arrhythmia present Check Patient and notify
Physician.
Blood pressure is out
of range.
Check Patient and verify BP
with manual method.
Improper cuff size / brand Measure patient limb. Use
only properly sized
accessories.
Reading too high or too low Incorrect cuff size Measure Patient limb, use
correct cuff.
Patient movement Wait until patient is calm or
gently hold limb.