Spectrum
®
/Spectrum OR
™
Service Manual 0070-10-0556-02 1 - 31
Repair Information Troubleshooting Menus
Sampling Error Appears when a sampling error
occurs on one or more Gas Module
channels during calibration
Repeat calibration
procedure. If problem
persists, contact Customer
Support.
Not Ready For
Calibration
Appears when the Gas Module is
unable to initialize calibration
Repeat calibration
procedure. If problem
persists, contact Customer
Support.
Calibration Error,
Sampling Error
Appears when a sampling error
occurs in all four Gas Module
channels during calibration
Repeat calibration
procedure. If problem
persists, contact Customer
Support.
Calibration Error, Zeroing
Error
Appears when the Gas Module
cannot perform a Zeroing during
calibration
Repeat calibration
procedure. If problem
persists, contact Customer
Support.
MESSAGE/PROBLEM REASON SOLUTION