Service Manual 0070-10-0556-02 1 - 39
Repair Information Installation Menu
The following table describes the Installation Menu structure:
MENU TITLE ON
SCREEN MENU CHOICES DEFAULT
ACTIONS/
COMMENTS
Save Current Select to save current
settings as defaults
Select Language Set up at factory Select to change
language
Select Country Set up at factory Select to change
country
Date Format M/D/Y, D/Y/M Per country Select to change date
format
Time Format 12, 24 hour Per country Select to change time
format
NIBP Timeout 15, 30, 45, 60 mins 15 Selects time old and
data is removed from
screen
CO & PAWP Timeout 15, 30, 45 mins, 1, 2,
4 hr
15 min Select to change CO
and PAWP timeout.
Enable EPM Yes, No No Select to enable EPM
module
Apnea Latch On, Off On Select to turn apnea
alarm latching on or
off.
Enable Mute All
Permanent Selection
Yes, No Yes Select to enable or
disable the Permanent
Audio Off menu
selection.
Enable Arrhythmia All
Off Selection
Yes, No No Select to enable or
disable the Arrhythmia
All Off menu selection.
Temperature Units ° F, ° C ° F - USA
° C - All others
Select to change
temperature units
Weight Units lbs, kg lbs - USA
kg - All others
Select to change weight
units
Height Units ft/ inches, cm ft/ inches - USA
cm - All others
Select to change height
units
CO
2
Units mmHg, %, kPa mmHg Select to change CO
2
units
SpO
2
Tones Standard Tones,
Alternate Tones
Standard Tones Select to change the
SpO
2
tones.
Copy monitor defaults
to card.
Select to copy the
monitor defaults and
settings to a data
transfer card inserted
into PCM2
Copy monitor defaults
from card.
Select to copy the
monitor defaults and
settings from a data
transfer card inserted
into PCM2

Table of Contents