13 - 26 Operator’s Manual
13 Probes and Biopsy
c. Check manually to confirm that the needle-guided bracket is securely installed on the
probe.
2. Install the guiding block:
a. Select a proper guiding block and push it into the groove above the angle block, and
clamp it tightly.
b. Insert a biopsy needle with the same specification as that of the guiding block into the
hole of the guiding block.
3. Release the needle from the bracket:
a. Remove the guiding block slightly along the direction of the needle’s tail.