F - 4 Operator’s Manual
F EMC Guidance and Manufacturer’s Declaration
Table F-2
GUIDANCE AND MINDRAY DECLARATION-ELECTROMAGNETIC EMISSIONS
The Probe P7-3Ts/P8-3Ts/P8-2Ts/LAP13-4Cs which integrated in the system (not including air
charge module) is intended for use in the electromagnetic environment specified below. The
customer or the user of system should assure that it is used in such an environment.
EMISSIONS TEST COMPLIANCE ELECTROMAGNETIC ENVIROMENT -
GUIDANCE
RF emissions
CISPR 11
Group 1 The Probes uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR
11
Class A The system is suitable for use in all
establishments other than domestic and those not
directly connected to the public low-voltage
power supply network of domestic residence.
WARNING
This device is not intended for
residential use and does not provide
adequate protection for radio reception.
RF interference caused by Class A equipment in a
specific environment.
When connecting Class A equipment to a low-
voltage power supply network, note that such
equipment is powered by a dedicated power
transformer or generator instead of a low-voltage
overhead power line;
During installation, take measures to reduce the
emission of Class A equipment, use special
power transformers or generators to supply power
instead of LV overhead power lines, and use
devices more than 30 m away from the residential
environment or a structure as a radiation barrier.
NOTE:
The EMISSIONS characteristics of this
equipment make it suitable for use in
industrial areas and hospitals (CISPR 11 class
A). If it is used in a residential environment
(for which CISPR 11 class B is normally
required) this equipment might not offer
adequate protection to radio-frequency
communication services. The user might need
to take mitigation measures, such as relocating
or re-orienting the equipment.
Harmonic Emissions
IEC 61000-3-2
--
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
--

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals