F EMC Guidance and Manufacturer’s Declaration
Operator’s Manual F - 5
GUIDANCE AND MINDRAY DECLARATION-ELECTROMAGNETIC EMISSIONS
The system not configured with air charge module or the probe P7-3Ts/P8-3Ts/P8-2Ts/LAP13-
4Cs is intended for use in the electromagnetic environment specified below. The customer or the
user of system should assure that it is used in such an environment.
EMISSIONS TEST COMPLIANCE ELECTROMAGNETIC ENVIROMENT -
GUIDANCE
RF emissions
CISPR 11
Group 1 The Probes uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR
11
Class B
The system is suitable for use in all
establishments including domestic establishments
and those directly connected to the public
lowvoltage power supply network that supplies
buildings used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Compliance
Table F-3
GUIDANCE AND MINDRAY DECLARATION-ELECTROMAGNETIC IMMUNITY
The system is intended for use in the electromagnetic environment specified below. The customer
or the user of system should assure that it is used in such an environment.
IMMUNITY TEST IEC 60601
TEST LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNET
IC ENVIRONMENT-
GUIDANCE
Electrostatic Discharge
(ESD)
IEC 61000-4-2
±8 kV contact;
±15 kV air
±8 kV contact;
±15kV air
Floors should be wood,
concrete or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Electrical fast
Transient / burst
IEC 61000-4-4
±2 kV for power
supply lines;
±1 kV for input/output
lines
±2 kV for power
supply lines;
±1 kV for input/output
lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s);
±2 kV line(s) to earth
±1 kV line(s) to line(s);
±2 kV line(s) to earth
Mains power quality
should be that of a
typical commercial or
hospital environment.

Table of Contents

Other manuals for Mindray TEX20

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray TEX20 and is the answer not in the manual?

Mindray TEX20 Specifications

General IconGeneral
BrandMindray
ModelTEX20
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals