6. Operating | 6.14 - Calibrating internal weighing point | 105
6.14.3 Modifying a adjustment
[Modify] may be used only for minor changes (e.g. changing the dead load,
adapting mV/V values for dead load and/or Max). Otherwise, always use
[New]!
Requirements:
Overwrite protection is deactivated, see ChapterCAL switch.
: [Adjustment]
1. Click [Modify] in the task bar.
2. Setting max. capacity [Max. capacity], see ChapterDetermining max.
capacity (Max).
3. Determining the dead load [Deadload], see ChapterDetermining the
dead load.
4. Activate the overwrite protection after finishing the adjustment, see
Chapter CAL switch.
6.14.3.1 Subsequent dead load correction
If the vessel/platform weight changes by an amount that is higher than the
zero range; e.g. due to dead load reduction, dead load increase, or mechan
ical changes, the functions for automatic zerotrack and manual zero setting
no longer work.
If the full zero-setting range is already being utilized, you can still correct the
dead load without affecting other adjustment data/parameters.
Requirements:
Overwrite protection is deactivated, see ChapterCAL switch.
The scale is unloaded.
1. Click [Modify adjustment] in the task bar.
2. Click [Modify dead load] in the task bar.
3. Click [by load] in the task bar.
4. Clear the scale/hopper.
5. Click [Set dead load] in the task bar to save the entries.
6. Save adjustment.

Table of Contents